| Model Number |
100-9810 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient implanted with a superion device crashed a golf cart and fell.During a steroid injection a broken cam lobe on the spacer was detected.The patient was not experiencing any symptoms caused by the broken cam lobe, and the device remains implanted.
|
| |
|
Search Alerts/Recalls
|