• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH GAMMA4 - RC LAG SCREW D10.5X85MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH GAMMA4 - RC LAG SCREW D10.5X85MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 81660085S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Necrosis (1971); Perforation (2001)
Date of Event 09/08/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information has been requested and if received, will be provided in the supplemental report.
 
Event or Problem Description
As reported: "the lag screw cut out.The patient had been following instructions.However, this time it was traumatic avascular necrosis caused by a fall.The event was discovered on (b)(6) 2025 and the nail was removed and an artificial head was implemented on (b)(6) 2025.".
 
Additional Manufacturer Narrative
The device has not been returned.The lag screw cut out could be confirmed by x-ray image received.The images show that the femoral had and neck collapsed over the system.Based on investigation, the root cause was attributed to a patient related issue.The patient fall led to a avascular necrosis, which ultimately caused the lag screw cut out.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Event or Problem Description
As reported: "the lag screw cut out.The patient had been following instructions.However, this time it was traumatic avascular necrosis caused by a fall.The event was discovered on (b)(6) 2025 and the nail was removed and an artificial head was implemented on (b)(6) 2025.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GAMMA4 - RC LAG SCREW D10.5X85MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key23219830
Report Number0008031020-2025-01667
Device Sequence Number9840493
Product Code HSB
UDI-Device Identifier07613327591798
UDI-Public(01)07613327591798(17)340430(10)K1B254A
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K230512
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number81660085S
Device Lot NumberK1B254A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/09/2025
Supplement Date Received by Manufacturer11/26/2025
Initial Report FDA Received Date10/06/2025
Supplement Report FDA Received Date12/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-