ETHICON INC. SFX SPI MCRYL+UNI VIO 6IN 3-0 SA RB-1 MP; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
|
Back to Search Results |
|
| Catalog Number |
SXMP1B449 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a hernia repair procedure on (b)(6) 2025 and barbed suture was used.During the procedure, it was reported that the suture was broken when the surgeon attempted to sew the tissue.Changed another one to complete the surgery.There is no report on patient's injury.No additional information could be provided.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6, d9, h3 investigation summary: the product received for analysis was sxmp1b449 visual inspection and metallurgical analysis were performed on the returned product.A failed suture needle, labeled as (b)(4), was submitted to (b)(6) for fractographic evaluation on october 20, 2025.The component was identified as product code sxmp1b449.It was requested that hsa assess the fracture mode of the failure.A fracture was observed at the suture attachment of the needle.The needle was received in two pieces, only one of the mating fracture surfaces was examined for this evaluation.A scanning electron microscope (sem) was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation of (b)(4) revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Macroscopic plastic deformation observed coincidental to the fracture provide additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.Corrected data: d4 udi.
|
| |
|
Search Alerts/Recalls
|
|
|