• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ALLIANCE GLENOID SIZER HANDLE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ALLIANCE GLENOID SIZER HANDLE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED Back to Search Results
Catalog Number SBGL2000
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).D10: item# sbgl3007; lot# unknown.G2: foreign - event occurred in canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that on an unknown date, it was discovered outside of a surgery that the center post reamer and sizer handle were both damaged/fractured.No patient involvement occurred.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d2; d4; d9; g1; g3; g6; h1; h2; h4; h10 d10: item# sbgl3007; lot# 65605176 if any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d2; g1; g3; g6; h1; h2; h3; h6.Visual examination of the returned products identified the reamer tip is fractured off on item sbgl3007 and the plunger is fractured on item sbgl2000.Fracture analysis has been previously analyzed for these fracture locations where overload was identified as the mode of failure.Both devices show signs of repeated use.Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.Reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.Complaint is confirmed.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLIANCE GLENOID SIZER HANDLE
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23245085
Report Number0001822565-2025-03685
Device Sequence Number13365655
Product Code KWS
UDI-Device Identifier00889024574632
UDI-Public(01)00889024574632(10)65316917
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K191814
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSBGL2000
Device Lot Number65316917
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2025
Supplement Date Received by Manufacturer10/08/2025
02/05/2026
Initial Report FDA Received Date10/08/2025
Supplement Report FDA Received Date10/17/2025
02/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NARRATIVE IN H11.
Patient SexUnknown
-
-