• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S TITAN INFLATABLE PENILE PROSTHESIS FAMILY; PENILE PROSTHESIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN INFLATABLE PENILE PROSTHESIS FAMILY; PENILE PROSTHESIS, INFLATABLE Back to Search Results
Catalog Number ESR820
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2025
Type of Reportable Event Serious Injury
Event or Problem Description
According to the available information the device was explanted and replaced as the device had experienced a mechanical failure and had no fluid in the system.
 
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
 
Event or Problem Description
According to the available information the device was explanted and replaced as the device had experienced a mechanical failure and had no fluid in the system.
 
Additional Manufacturer Narrative
Titan otr pump, cylinders 1 and 2, and reservoir were received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.No functional abnormalities were noted with the pump.Abrasion was noted on the exhaust tubes cylinder 1 and cylinder 2.A separation, surrounded by abrasion, was noted on the exhaust tube of cylinder 1.This is a site of leakage.See attached photo.No functional abnormalities were noted with cylinder 2.Abrasion was noted on the inlet tube of the reservoir.No functional abnormalities were noted with the reservoir.Based on examination of the returned product, it was concluded that while in-vivo both the exhaust tubes and inlet tube of the pump had overlapped and abraded against one another.This positioning, in combination with device usage over time, could contribute to sufficient stress to separate the exhaust tubing of cylinder 1.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformance's and capas revealed no trends for this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN INFLATABLE PENILE PROSTHESIS FAMILY
Common Device Name
PENILE PROSTHESIS, INFLATABLE
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer (Section G)
MINNEAPOLIS
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key23247785
Report Number2125050-2025-01627
Device Sequence Number18703650
Product Code FHW
UDI-Device Identifier05708932487566
UDI-Public5708932487566
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P000006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/18/2017
Device Catalogue NumberESR820
Device Lot Number3739908_ESR8202400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/08/2025
Supplement Report FDA Received Date12/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/19/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexUnknown
-
-