|
|
| Catalog Number |
391.962 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cutting slot of the pliers are broken.A small piece fell off.A replacement cutter was used.The procedure was successfully completed with no surgical delay.There was no patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d4 (lot, expiration date), d9, h4 corrected: d4 primary udi number the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: corrected: g1.H3, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that cut area of the bend/cutt pliers was broken and corrosion was noted on that.The fragment was not received.No other issues were identified.A dimensional inspection for thebend/cutt pliers was not performed due to post manufqacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.Properly handling and attention to the approved use of the device diminishes the risk of failure.Corrosion: the potential cause is being attributed to cleaning/sterlization methods.As per se_023827 rev ap, during cleaning, only use cleaning agents that are labelled for use on medical devices and in accordance with the manufacturer¿s instructions (e.G.Temperature, contact time, and rinse time).Cleaning agents with a used dilution ph of within 7¿9.5 are recommended.Highly alkaline conditions (ph >11) can damage components/devices, such as aluminum materials.Do not use saline, environmental disinfection (including chlorine solutions) or surgical antiseptics (such as iodine- or chlorhexidine-containing products).Do not use a cleaning aid that can damage the surface of instruments such as steel wool, abrasive cleaners or wire brushes.A functional test was not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the bend/cutt pliers would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.There is no indication that a design or manufacturing issue has caused the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot = > part number: 391.962.Lot number: t147153.Product code: n/a.Date of manufacture: 19-may-2017.Manufacturing location: tuttlingen.Part expiration date: n/a.Nonconformance noted: n/a.Dhr record review: a manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.The raw material certificate was reviewed and the used material was according to the specification of the device.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|