Catalog Number 03-2722-9
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 06/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A user facility nurse manager (nm) reported that a combi set blood leak occurred approximately eighty-four minutes into a patient¿s hemodialysis (hd) treatment.There were two small drops of blood noted on the base of the machine.The blood leak was coming from the portion of the tubing that is placed inside the blood pump.There was a small pinhole observed at the center of the tubing.Blood from the arterial line was returned.Blood from the venous line could not be returned.The patient's estimated blood loss was 200ml.There were no machine alarms.No damage or defects were noted on the lines prior to use, and no loose connections were noticed.The nm confirmed there were no leaks during the priming phase.There were no changes or disruptions in pressure that could have contributed to the leak.The patient did not experience any adverse effects or require medical intervention due to the blood loss.The patient was re-setup with new supplies and completed their treatment on the same machine.The blood pump was inspected following the event and no sharp edges were noted.The combi set was not available to be returned for evaluation.
 
Event or Problem Description
A user facility nurse manager (nm) reported that a combi set blood leak occurred approximately eighty-four minutes into a patient¿s hemodialysis (hd) treatment.There were two small drops of blood noted on the base of the machine.The blood leak was coming from the portion of the tubing that is placed inside the blood pump.There was a small pinhole observed at the center of the tubing.Blood from the arterial line was returned.Blood from the venous line could not be returned.The patient's estimated blood loss was 200ml.There were no machine alarms.No damage or defects were noted on the lines prior to use, and no loose connections were noticed.The nm confirmed there were no leaks during the priming phase.There were no changes or disruptions in pressure that could have contributed to the leak.The patient did not experience any adverse effects or require medical intervention due to the blood loss.The patient was re-setup with new supplies and completed their treatment on the same machine.The blood pump was inspected following the event and no sharp edges were noted.The combi set was not available to be returned for evaluation.
 
Additional Manufacturer Narrative
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the facility for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius combi sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM COMBI SET
Common Device Name
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key23251435
Report Number0008030665-2025-02409
Device Sequence Number13323493
Product Code FJK
UDI-Device Identifier00840861100293
UDI-Public00840861100293
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K962081
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 11/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03-2722-9
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2025
Supplement Date Received by Manufacturer11/25/2025
Initial Report FDA Received Date10/08/2025
Supplement Report FDA Received Date11/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS 2008T MACHINE; FRESENIUS 2008T MACHINE; FRESENIUS OPTIFLUX DIALYZER; FRESENIUS OPTIFLUX DIALYZER
Patient Age56 YR
Patient SexFemale
Patient Weight63 KG
-
-