Catalog Number 115-435
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/15/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Patient reported left side "fluid around implant and small leak" (captured as rupture).Later healthcare professional reported "textured recalled implants".This was diagnosed via mri.The device remains implanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b5, b6, d6b, h6.
 
Event or Problem Description
Device has been explanted and replaced.
 
Event or Problem Description
Patient reported left side "fluid around implant and small leak" (captured as rupture).Later healthcare professional reported "textured recalled implants".This was diagnosed via mri.The device remains implanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b6.
 
Additional Manufacturer Narrative
Device evaluation: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: observed broken device through microscopic inspection assessed as unidentified (tear) opening.No further actions are required as it is not related to the manufacturing process.¿ anxiety-product/procedure: unable to observe through visual inspection as it is a physiological phenomenon.No additional observations performed on the device.No further actions are required.
 
Event or Problem Description
Patient reported left side "fluid around implant and small leak" (captured as rupture).Later healthcare professional reported "textured recalled implants".This was diagnosed via mri.The device remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 115 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key23256979
Report Number9617229-2025-17258
Device Sequence Number10236482
Product Code FTR
UDI-Device Identifier10888628002425
UDI-Public10888628002425
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/16/2018
Device Catalogue Number115-435
Device Lot Number2419631
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/17/2025
Supplement Date Received by Manufacturer12/09/2025
01/20/2026
02/26/2026
Initial Report FDA Received Date10/09/2025
Supplement Report FDA Received Date12/29/2025
02/06/2026
03/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/16/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
Patient Weight61 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-