(b)(4).The sample was returned to the manufacturer for evaluation.The manufacturer reported: "per customer provided information the dhr for the alleged defective medical device was reviewed and found complete without any irregularities.The alleged defective instrument was produced at the tecomet, inc.Kenosha, wl facility as part of a (b)(4) pc.Lot in november of 2024 from lot number 06e2467427 ((b)(6)).The investigation determined that the manufacturing processes adhered to approved specifications, utilizing the correct materials and components.A thorough visual and functional inspection was performed, which revealed that the jaws are entirely missing from the outer tube.This failure led to the jaws becoming fully detached from the device.While the root cause of the detachment and rivet dislodgement could not be conclusively determined, it is suspected that excessive force applied during use at the end user's facility contributed to the failure.Notably, all (b)(4) instruments from this lot underwent 100% visual inspection and functional testing prior to shipment, in accordance with standard operating procedures for this product line.Due to these findings, no further actions will be taken in response to this complaint, and this record will be deemed closed." teleflex will continue to monitor and trend on complaints of this nature.Other remarks: n/a corrected data: n/a.
|