| Catalog Number |
0165SI20 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter facility name: (b)(6) hospital.The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter connector for injecting sterile water had come off when the product was opened.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed ¿ unknown cause.Evaluation confirmed that the detached valve and cap on the returned catheter.Further investigation was documented under cid#(b)(4).A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Therefore, no additional action is required at this time.Corrections made to tab(s) d and h.Initial reporter complete facility name: (b)(6).Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter connector for injecting sterile water had come off when the product was opened.
|
| |
|
Search Alerts/Recalls
|