| Catalog Number |
VCP310H |
| Medical Device Problem Codes |
Material Separation (1562); Defective Component (2292); Dull, Blunt (2407); Structural Problem (2506); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.The following was reported: "the suture came off the needle.The surgeon noted that this needle felt thinner and less durable than those previously used, lacking sharpness and stability." quantity of product involved was entered as (b)(4).* please confirm the number of devices in which the needle came off the needle during this procedure.* what is the procedure name and date? * please confirm the number of devices in which "this needle felt thinner and less durable than those previously used, lacking sharpness and stability"? were (b)(4) devices affected by this issue? * please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.* please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep or loc/bq).Additional h11: was surgery delayed due to the reported event? --> unknown.Was procedure successfully completed? --> unknown.Were fragments generated? --> unknown if yes, were they removed easily without additional intervention? --> unknown.Patient status/ outcome / consequences --> no.Was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: --> unknown.Is the patient part of a clinical study --> unknown.(b)(4).Device property of -->none.Device in possession of -->none.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure in 2025 and suture was used.During the procedure, the suture came off the needle.The surgeon noted that the needle was not as sharp as those from previous batches, felt thinner, and seemed weaker and less stable.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected data: d4 udi.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Date sent to the fda: 1/16/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 pending evaluation of returned device additional information: d9, h3, h6 a device has been received, however it has not yet been evaluated.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 3/2/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigational summary: the product was returned to ethicon for evaluation.One sealed box with thirty-six representative unopened samples was received for analysis.Product code vcp310h.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damage was evident on the exterior packaging.Following the sampling plan, a visual inspection was performed on thirteen samples.Packets were opened and the needles removed from their trays.The swage and attachment areas were consistent with expectations, and all needles conformed to the specification for product code vcp310h (taper point, sales type sh-1 plus, ½ circle, needle size 18 mils).No bending, tip damage, or deformation was observed during inspection.Needle diameters were measured and met the requirements.Needles were also passed through a tissue simulator with no abnormalities noted.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no anomalies were noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
|
| |
|