| Catalog Number |
71800110 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Fall (1848)
|
| Date of Event |
09/13/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, on (b)(6)2025, the patient sustained a fracture of the right distal femur from the distal 3rd junction, extending into the articular area due to a fall.An orif was performed on (b)(6)2025 and a 4.5mm l-d fem lk pl10h r 230mm was implanted for fracture fixation, along with two cables.After this, the patient presented again on (b)(6)2025 with a history of falling.The x-ray showed that the 4.5mm l-d fem lk pl10h r 230mm broke on the 8th hole from proximal.A revision surgery was performed on (b)(6)2025 to remove the broken plate, insert a retrograde nail and supplement fracture with a competitor distal femur peri prosthetic plate.The current health status of the patient is unknown.
|
| |
|
Search Alerts/Recalls
|
|
|