MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TAPERLOC COMPLETE PRIMARY FEMORAL 133 T1 PPS SO 18X156MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TAPERLOC COMPLETE PRIMARY FEMORAL 133 T1 PPS SO 18X156MM; PROSTHESIS, HIP Back to Search Results
Catalog Number 51-103180
Medical Device Problem Codes Tear, Rip or Hole in Device Packaging (2385); Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the packaging was not sealed and there was a small piece of debris inside.There was no known impact or consequences to the patient.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected}.Updated: g3; h2; h3; h6.Evaluation of the photograph provided confirmed there is white debris inside the sterile packaging which is visually consistent with the appearance of foam debris from the foam packaging and damage to the sterile packaging (pouch).Sterility has been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERLOC COMPLETE PRIMARY FEMORAL 133 T1 PPS SO 18X156MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23277292
Report Number0001825034-2025-03178
Device Sequence Number5560035
Product Code LZO
UDI-Device Identifier00880304498358
UDI-Public(01)00880304498358(17)320816(10)7309203
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K200196
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number51-103180
Device Lot Number7309203
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2025
Supplement Date Received by Manufacturer02/03/2026
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date02/04/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-