• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ACTIS BROACH SZ 10; BROACHES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ACTIS BROACH SZ 10; BROACHES Back to Search Results
Catalog Number 201001100
Medical Device Problem Codes Break (1069); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that actis broaches need to be replaced.There were severe stress rings on them where they attach to broach handle.Surgeon frequently breaks them.When they break it is extremely hard to remove from the patient.(b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d4 (lot), d9, h4.Corrected: d3, d4 (primary udi #).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> it was reported that actis broaches needed to be replaced.There were severe stress rings on them where they attach to broach handle.Surgeon frequently breaks them.When they break it is extremely hard to remove from the patient.This d.E.R.Is preventitive.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device revealed that the actis broach had signs of extensive use.Wear was observed on the coupling area, which is an indicator of repeated use.A functional test was performed with a laboratory sample mating device, and the actis broach was successfully assembled without any issues.There were no signs of looseness or difficulties observed during assembly.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the actis broach would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS BROACH SZ 10
Common Device Name
BROACHES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork IN
EI 
Manufacturer Contact
elba bello
700 orthopaedic drive
warsaw, IN 08869
9083863534
MDR Report Key23277807
Report Number1818910-2025-17452
Device Sequence Number8809291
Product Code HTQ
UDI-Device Identifier10603295393726
UDI-Public(01)10603295393726(10)PG314157
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K150862
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number201001100
Device Lot NumberPG314157
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/22/2025
Supplement Date Received by Manufacturer10/16/2025
10/20/2025
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date10/20/2025
10/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age77 YR
Patient SexFemale
-
-