MAUDE Adverse Event Report: BIOTRONIK AG PK PAPYRUS US 2.5/15; COVERED CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG PK PAPYRUS US 2.5/15; COVERED CORONARY STENT SYSTEM Back to Search Results
Model Number 434887
Medical Device Problem Codes Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
A pk papyrus covered stent system was selected for treatment of a type ii perforation in the mid rca.The device could not be delivered and then the stent got lost.The stent could not be found, neither inside nor outside patient's body.Another pk papyrus was implanted in an additional intervention.The physician rated the event as device related complication.
 
Event or Problem Description
A pk papyrus covered stent system was selected for treatment of a type ii perforation in the mid rca.The device could not be delivered and then the stent got lost.The stent could not be found, neither inside nor outside patient's body.Another pk papyrus was implanted in an additional intervention.The physician rated the event as device related complication.
 
Additional Manufacturer Narrative
The affected pk papyrus stent system was not returned and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form, cathlab report) and the production documentation were reviewed to establish whether a device deficiency contributed to this event.Review of the production documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause could be identified.The treating physician rated the occurrence as no procedure- but device-related complication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
 
Additional Manufacturer Narrative
The affected pk papyrus stent system was not returned and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form, cathlab report) and the production documentation were reviewed to establish whether a device deficiency contributed to this event.Review of the production documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause could be identified.The treating physician rated the occurrence as no procedure- but device-related complication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.Additional information (10 dez 2025): product data has been upated.
 
Event or Problem Description
A pk papyrus covered stent system was selected for treatment of a type ii perforation in the mid rca.The device could not be delivered and then the stent got lost.The stent could not be found, neither inside nor outside patient's body.Another pk papyrus was implanted in an additional intervention.The physician rated the event as device related complication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PK PAPYRUS US 2.5/15
Common Device Name
COVERED CORONARY STENT SYSTEM
Manufacturer (Section D)
BIOTRONIK AG
ackerstrasse 6
bulach, zurich 8180
SZ  8180
Manufacturer (Section G)
BIOTRONIK AG
ackerstrasse 6
bulach, zurich 8180
SZ   8180
Manufacturer Contact
tina mueller
ackerstrasse 6
bulach, zurich 8180
SZ   8180
MDR Report Key23277975
Report Number8043892-2025-09302
Device Sequence Number20213817
Product Code NIV
UDI-Device Identifier07640130447158
UDI-Public(01)07640130447158(17)270409(10)03251458
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
H17004
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number434887
Device Lot Number03251458
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2025
Supplement Date Received by Manufacturer09/30/2025
09/30/2025
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date11/11/2025
12/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/15/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-