| Model Number |
434887 |
| Medical Device Problem Codes |
Failure to Advance (2524); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A pk papyrus covered stent system was selected for treatment of a type ii perforation in the mid rca.The device could not be delivered and then the stent got lost.The stent could not be found, neither inside nor outside patient's body.Another pk papyrus was implanted in an additional intervention.The physician rated the event as device related complication.
|
| |
|
Event or Problem Description
|
|
A pk papyrus covered stent system was selected for treatment of a type ii perforation in the mid rca.The device could not be delivered and then the stent got lost.The stent could not be found, neither inside nor outside patient's body.Another pk papyrus was implanted in an additional intervention.The physician rated the event as device related complication.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected pk papyrus stent system was not returned and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form, cathlab report) and the production documentation were reviewed to establish whether a device deficiency contributed to this event.Review of the production documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause could be identified.The treating physician rated the occurrence as no procedure- but device-related complication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected pk papyrus stent system was not returned and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form, cathlab report) and the production documentation were reviewed to establish whether a device deficiency contributed to this event.Review of the production documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause could be identified.The treating physician rated the occurrence as no procedure- but device-related complication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.Additional information (10 dez 2025): product data has been upated.
|
| |
|
Event or Problem Description
|
|
A pk papyrus covered stent system was selected for treatment of a type ii perforation in the mid rca.The device could not be delivered and then the stent got lost.The stent could not be found, neither inside nor outside patient's body.Another pk papyrus was implanted in an additional intervention.The physician rated the event as device related complication.
|
| |
|
Search Alerts/Recalls
|