• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. SAW BLADES LINVATEC®/HALL® POWER SYSTEMS; SURGICAL SAW BLADE, OSCILLATING, SINGLE-USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. SAW BLADES LINVATEC®/HALL® POWER SYSTEMS; SURGICAL SAW BLADE, OSCILLATING, SINGLE-USE Back to Search Results
Catalog Number 11-3583
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during inspection the device had a hair-like substance inside the packaging.There was no patient involvement.Due diligence is complete and there is no additional information available.
 
Additional Manufacturer Narrative
An investigation into the reported event has been initiated under (b)(4).Once the investigation has been completed, a follow up report will be submitted with the investigation results and any actions taken by the manufacturer.E1 telephone number: (b)(6).G2 country: japan.
 
Additional Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d2, d4, d9, e2, g1, g3, g6, h1, h2, h3, h4, h5, h6, and h11.Package of blade was returned unopened.Visual inspection did find a hair-like debris sealed inside the packages.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The reported event was confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
There is no additional information regarding the event.
 
Additional Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).Following sections were updated or corrected: b4, b5, d2, g1, g3, g6, h1, h2, h10, and h11.This additional report is being sent to add related report number.
 
Event or Problem Description
There is no additional information regarding the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAW BLADES LINVATEC®/HALL® POWER SYSTEMS
Common Device Name
SURGICAL SAW BLADE, OSCILLATING, SINGLE-USE
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key23278365
Report Number0001526350-2025-01308
Device Sequence Number8853823
Product Code GFA
UDI-Device Identifier00889024494015
UDI-Public(01)00889024494015(17)300729(10)476943
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number11-3583
Device Lot Number476943
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2025
Supplement Date Received by Manufacturer04/21/2026
04/21/2026
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date04/24/2026
04/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-