H11: additional manufacturer narrative: the device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
It was reported that a patient with a 29mm 7300tfx mitral valve underwent a valve-in-valve procedure after an implant duration of 5 years, 7 months due to stenosis and regurgitation.The patient presented with symptoms of heart failure.The tmvr was performed with a 29mm 9755rsl transcatheter valve.The patient tolerated the procedure well, was in stable condition post procedure, and discharged on pod #2.
|
|
H11: additional manufacturer narrative: updated: b4, g3, g6, h2, h6 (type of investigation, investigation findings, investigation conclusions).Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.There is insufficient information available regarding patient or procedural factors known to cause or contribute to stenosis.Therefore, a definitive root cause cannot be conclusively determined.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.
|