• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION THE VEST SYSTEM; PERCUSSOR, POWERED-ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION THE VEST SYSTEM; PERCUSSOR, POWERED-ELECTRIC Back to Search Results
Model Number P105
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The vest airway clearance system, model 105 was developed to provide effective airway clearance therapy.The system consists of an inflatable garment attached to an air pulse generator that rapidly inflates and deflates the inflatable garment.This causes the chest wall to be gently compressed and released, which creates airflow within the lungs.This process moves the mucus toward the large airways where it can be cleared by coughing or suctioning.This type of airway clearance therapy is referred to as high frequency chest wall oscillation (hfcwo).Customer confirmed that the device was discarded.Product was not available and a solution could not be provided.Therefore, based on this information, no further action is required.Although there was no reported injury with this event, if the report of a damaged power cord with exposed wires were to recur, it could potentially cause serious injury or death.Therefore, baxter is reporting this event.
 
Event or Problem Description
Baxter received a report stating that vest model 105 power cord had damage with exposed copper wires.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THE VEST SYSTEM
Common Device Name
PERCUSSOR, POWERED-ELECTRIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
HILL-ROM SINGAPORE
1 yishun avenue 7
singapore, north east 76892 3
SN   768923
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key23280521
Report Number3008145987-2025-00014
Device Sequence Number12952248
Product Code FYI
UDI-Device Identifier00887761997776
UDI-Public(01)00887761997776
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberP105
Device Catalogue NumberP105
Device Lot NumberN/A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2025
Initial Report FDA Received Date10/15/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NI.
Patient SexUnknown
-
-