ENCORE MEDICAL L.P. DJO SURGICAL; DJO EMPOWR KNEETM, PRESS FIT BP MINUS, 7R Back to Search Results
Catalog Number 342-16-707
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Codes Joint Laxity (4526); Implant Pain (4561)
Date of Event 09/16/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient required revision surgery after experiencing pain and instability.The right 7 minus porous baseplate had failed.Surgeon revised the construct using a 7 right minus universal baseplate, a 12x65 mm stem, two 7x5 mm augment blocks (one on the lateral side and one on the medial side), one 6x5 mm augment block on the lateral side, and a 7x16 mm 3d insert.
 
Additional Manufacturer Narrative
Once the investigation is completed, any additional information will be reported in a supplemental report.
 
Additional Manufacturer Narrative
See d2, d4 expiration date, d10, h4, h6, h10 and h11.The agent reported "(patient had complained of pain & instability.Patient had complained of pain & instability.The right 7 minus porous baseplate had failed.Doctor replaced that with a 7 right minus universal baseplate,12x65mm stem, (2) 7 x5mm augment blocks one on lateral side one on medial side, (1) 6x5mm augment block on lateral side, and used a 7x16mm 3d insert.)".The previous surgery and the surgery detailed in this event occurred 2 months and 2 weeks apart.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.The root cause of this complaint was a revision surgery due to pain, instability and failed baseplate.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, patient activities or trauma.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
DJO EMPOWR KNEETM, PRESS FIT BP MINUS, 7R
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd.
austin, TX 78758
MDR Report Key23286218
Report Number1644408-2025-01493
Device Sequence Number5559406
Product Code MBH
UDI-Device Identifier00888912167642
UDI-Public00888912167642
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K143242
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number342-16-707
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2025
Supplement Date Received by Manufacturer11/26/2025
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date12/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/05/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
342-16-707 LOT: 161T1076; 353-04-107 LOT CODE UNKNOWN
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexFemale
-
-