| Catalog Number |
MZ9271 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd maxzero extension set was leaking.The following information was received by the initial reporter with the following: it was reported by customer that they have another leaking filter.
|
| |
|
Additional Manufacturer Narrative
|
|
One sample model mz9271 was returned for investigation.The set was examined for defects and abnormalities.No defects or abnormalities were observed.The set was flushed with water and a leak was coming from the vent verifying the customer complaint.Additional air under water test was performed.The set was pressurized with about 10psi and the vent was submerged under water.There was a steady stream of air bubbles coming from the filter vent.This would indicate that there is no issue with damage/ holes in the membrane.The reason for the membrane leaking is that the hydrophobicity of the membrane must have been compromised during use and not during the manufacturing process.A device history record review was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.
|
| |
|