| Catalog Number |
AC4340-01 |
| Medical Device Problem Codes |
Material Separation (1562); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/19/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, during a bankart repair, it was noticed that white plastic fell off from a covator tip.It was split into two pieces; one piece was retrieved using a grasper.Another piece was left at the bottom of the patient¿s shoulder as it was too difficult to reach; this could result in worn on the patient¿s shoulder.The procedure was performed, after a non-significant delay, using a similar s+n back-up product.Patient status is normal, recovering.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).H11 h3, h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Search Alerts/Recalls
|