SMITH & NEPHEW, INC. FOOTPRINT ULR 4.5 PK SUT SL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72203783
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/21/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during an acl reconstruction, a 4mm footprint drill was used to create a bone hole, upon the insertion, it was noticed that the distal tip of the footprint ultra anchor snapped.The device was removed with tweezers.The procedure was resumed, without any delay, using an equivalent sn back-up.The back up device was used in the same bone hole and no material was left in the patient.The insertion site was cleared of all debris prior to insertion of the anchor.No further complications were reported.
 
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOTPRINT ULR 4.5 PK SUT SL
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key23287061
Report Number1219602-2025-02554
Device Sequence Number17532695
Product Code MBI
UDI-Device Identifier00885554029406
UDI-Public885554029406
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K123579
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72203783
Device Lot Number2169754
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2025
Initial Report FDA Received Date10/15/2025
Date Device Manufactured11/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age14 YR
Patient SexUnknown
-
-