|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The interlock was overridden and the loading unit was applied to test media.All staples were placed and test media was cleanly transected.The loading unit interlock was tested and found to function properly.It was reported that during use, the instrument did not fire after grasping the tissue.A new blue reload was replaced with the same powered handle and adapter, and the case continued.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the instrument firing handle had been partially compressed and released after pressing the green firing button.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: in order to fire the instrument, push the green button.Squeeze the handle sequentially until the oval clamp cover reaches the distal end of the cartridge slot, and the handle locks.Sequential squeezes of the handle are required to fully fire the loading unit.The total number of squeezes is relative to the length of the loading unit (30, 45 or 60).Failure to completely fire the loading unit will result in an incomplete cut and/or incomplete staple formation, which may result in poor hemostasis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|