| Catalog Number |
123512A |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Initial reporter phone: (b)(6).This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use there was black foreign matter inside the foley catheter packaging.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use there was black foreign matter inside the foley catheter packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed - cause unknown.The product had caused the reported failure.Evaluated the sample and observed a loose black dirt outside the catheter polysleeve.However, the exact cause on how and when problem occurred could not be determined.A potential root cause for this failure mode could be, ¿foreign matter from unclean working area¿.A review of the device labeling has been conducted for investigation purposes and was found adequate.A review of the device history record did not show any problems or condition that would have contributed to the reported issue.Therefore, no additional action required at this time.Corrections: d, h.Initial reporter phone:(b)(6).This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|