| Model Number |
Mi1250 SYNCHRONY 2 |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
No coupling is possible to the implant.Parents say that the user start learning walking so it might be he had a head trauma over the implant site.Re-implantation is planned.
|
| |
|
Event or Problem Description
|
|
Parents say that the user start learning walking so it might be he had a head trauma over the implant site.Re-implantation is planned.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the information received from the field a technical implant failure seems likely.However, to determine an exact root cause device investigation of the explanted device would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
Parents say that the user start learning walking so it might be he had a head trauma over the implant site.The user has been reimplanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the device investigation revealed mechanical damage to the stimulator electronics that is typical for severe external impact to the housing.The problems described in the recipient report appear to match the damage found.This is a final report.
|
| |
|