• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SYNCHRONY; Cochlear Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SYNCHRONY; Cochlear Implant Back to Search Results
Model Number Mi1250 SYNCHRONY 2
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
No coupling is possible to the implant.Parents say that the user start learning walking so it might be he had a head trauma over the implant site.Re-implantation is planned.
 
Event or Problem Description
Parents say that the user start learning walking so it might be he had a head trauma over the implant site.Re-implantation is planned.
 
Additional Manufacturer Narrative
Additional information: according to the information received from the field a technical implant failure seems likely.However, to determine an exact root cause device investigation of the explanted device would be necessary.Re-implantation is considered but no date has been scheduled yet.
 
Event or Problem Description
Parents say that the user start learning walking so it might be he had a head trauma over the implant site.The user has been reimplanted.
 
Additional Manufacturer Narrative
Conclusion: the device investigation revealed mechanical damage to the stimulator electronics that is typical for severe external impact to the housing.The problems described in the recipient report appear to match the damage found.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Common Device Name
Cochlear Implant
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
+43()577885626
MDR Report Key23290184
Report Number9710014-2025-00950
Device Sequence Number18690336
Product Code MCM
UDI-Device Identifier09008737369026
UDI-Public(01)09008737369026
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/03/2023
Device Model NumberMi1250 SYNCHRONY 2
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/23/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2025
Supplement Date Received by Manufacturer10/12/2025
10/12/2025
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date11/03/2025
08/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/03/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age17 MO
Patient SexMale
-
-