|
(b)(4).Date sent: 10/14/2025.D4: batch #855a69.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the ecr45g reload was returned fully fired with the reload pan partially dislodged and damaged (scratches) from the left proximal side.In addition, 5 double driver were missing with staples in the pockets, making the reload non-functional.In addition, an opened package was received along with the reload.No functional test was performed due the condition of the reload.No conclusion could be reached on what may have caused the reload pan to dislodge.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device ecr45g with lot number 928a09; and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 855a69, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|