• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. RS EXPANDED GLENOSPHERE 42MM, +4MM OFFSET; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. RS EXPANDED GLENOSPHERE 42MM, +4MM OFFSET; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 320-02-42
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Muscle/Tendon Damage (4532)
Date of Event 07/05/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an initial reverse total shoulder replacement on the right side.Subsequently, the patient experienced the disassembly and migration of the polyethylene component.As a result, approximately 2 years after initial replacement, the patient underwent a right shoulder revision.No further impact to the patient was reported.No other information is available.
 
Additional Manufacturer Narrative
The revision reported was likely the result of disassembly between the humeral liner and humeral tray and subsequent migration of the humeral liner.The reported wear of the adapter tray and glenosphere was likely due to the migration of the liner and subsequent metal-on-metal articulation between the tray and glenosphere.The migration of the liner may have been the result of disassociation from the tray and internal forces in vivo.However, this cannot be confirmed and potential contributions from patient conditions, an unreported traumatic event, incomplete seating of the humeral liner during implantation, forceful contact between the liner and glenoid bone (scapular notching), or a combination of the above to the reported disassembly of the humeral liner cannot be determined as the devices were not returned for evaluation and images/radiographs were not provided.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RS EXPANDED GLENOSPHERE 42MM, +4MM OFFSET
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key23291268
Report Number1038671-2025-03107
Device Sequence Number20049177
Product Code KWT
UDI-Device Identifier10885862186768
UDI-Public10885862186768
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K110708
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Catalogue Number320-02-42
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2025
Initial Report FDA Received Date10/15/2025
Date Device Manufactured06/22/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight70 KG
-
-