NEW WORLD MEDICAL, INC. AHMED GLAUCOMA VALVE; GLAUCOMA DRAINAGE DEVICE Back to Search Results
Model Number FP7
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record for the reported lot was reviewed with no discrepancies identified.The device was manufactured, tested, and released per validated procedures.Although the initial report indicated the device was available for evaluation, repeated attempts to contact the customer have been unsuccessful.The reported issue appears related to drainage redirection, a normal valve function where fluid exits along the valve pocket and may resemble leakage during priming.The device has not yet been returned to new world medical.If received, it will be evaluated and an addendum to this report will be submitted.
 
Event or Problem Description
These items were reported to be defective by the surgeon.When irrigating both shunts, the fluid came from the top of the channel.Pt code: 4582.Device codes: 2588, 1354.Ref reports: mw5174471, mw5174472, mw5174473.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AHMED GLAUCOMA VALVE
Common Device Name
GLAUCOMA DRAINAGE DEVICE
Manufacturer (Section D)
NEW WORLD MEDICAL, INC.
10763 edison court
rancho cucamonga CA 91730
Manufacturer (Section G)
NEW WORLD MEDICAL, INC.
10763 edison court
rancho cucamonga CA 91730
Manufacturer Contact
cristina avalos
10763 edison court
rancho cucamonga, CA 91730
9094664304
MDR Report Key23292426
Report Number1000125279-2025-00034
Device Sequence Number9926542
Product Code KYF
UDI-Device Identifier00892064002119
UDI-Public(01)00892064002119(11)250113(17)270113(10)A0225(21)
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K230975
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 10/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFP7
Device Catalogue NumberFP7
Device Lot NumberA0225
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2025
Initial Report FDA Received Date10/15/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/13/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-