Model Number LOCK DISTAL FIBULA PLT, SS LT, 8H
Medical Device Problem Codes Positioning Failure (1158); Defective Device (2588)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 10/3/2025, it was reported by a sales representative via email that a 2.7 screw would not lock into an ar-8943dl-08 locking distal fibula plate.This was discovered during a procedure on (b)(6) 2025.Additional information was received on 10/8/2025: it was confirmed that the 2.7 screw did not lock into three ar-8943dl-08 locking distal fibula plate during the ankle fixation surgery.When the plates were inspected, they did not have a stamp like the other plates of different size.The case was completed using a 10 hole tubular plate.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.The complaint allegation is confirmed based on photographic evidence of an ar-8943dl-08, batch 105642503, that displays the screw was not locking into the designated plate.Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude the most likely cause.The most likely cause is an internal manufacturing issue, as the plate was not marked correctly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCK DISTAL FIBULA PLT, SS LT, 8H
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key23292520
Report Number1220246-2025-04582
Device Sequence Number14648369
Product Code HRS
UDI-Device Identifier00888867346772
UDI-Public00888867346772
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K123241
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 02/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLOCK DISTAL FIBULA PLT, SS LT, 8H
Device Catalogue NumberAR-8943DL-08
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date10/03/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2025
Supplement Date Received by Manufacturer01/29/2026
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date02/03/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-