A customer reported after removing the abbott diabetes care (adc) device from the ac adapter after approximately 1.5 hours of charging, the device came into contact with the area just below their left wrist, resulting in a burn and scar at the site.The customer experienced pain, discomfort and had contact with a health care professional (hcp) who prescribed an unspecified ointment for treatment.There was no report of death or permanent impairment associated with this event.
|
|
The product has been requested back for investigation and the customer agreed to return the device; however, at this time product has not yet been received.An investigation has been performed for the reported complaint, and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the product was reviewed and the dhrs showed the product passed all tests prior to release.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components or misuse.However, mitigation's are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.If product is returned, a physical investigation will be performed per adc's established processes and procedures and a report will be submitted upon completion of investigation.All pertinent information available to abbott diabetes care has been submitted.
|