| Model Number |
B1060-040 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure performed was to treat a restenosed, heavily calcified, heavily tortuous arteriovenous fistula artery that is 80% stenosed.The 6x40mm armada 35 balloon dilatation catheter (bdc) ruptured during the first inflation at 14 atmospheres.A new armada 35 bdc was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection and functional testing were performed on the returned device.The reported balloon rupture was not confirmed.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the reported information and the observations from the returned analysis, a definitive cause for the reported balloon rupture could not be determined.Possible factors that could contribute to balloon material ruptures include, but are not limited to, inflation technique, interactions with other devices or patient anatomy.In this case, a conclusive cause for the reported balloon rupture could not be determined since the complaint could not be confirmed during return analysis.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
Subsequently after the initial report had already been filed, device analysis indicates that there was no device malfunction as the balloon catheter was inflated and maintained pressure.No balloon rupture was noted.No additional information was provided.
|
| |
|