| Catalog Number |
10-210 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/02/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture" confirmed via mri.This record is for the right side.Device remains implanted.Device will be returned.
|
| |
|
Event or Problem Description
|
|
Device was explanted and replaced.Device will be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d6b, h6.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device related to the reported events rupture was received on nov 20, 2025, with lot number 2023051.Per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: observed an opening through microscopic inspection assessed as fold flaw opening and observed a missing piece of shell through microscopic inspection assessed as inconclusive.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis (non-penetrating nicks) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture" confirmed via mri.This record is for the right side.Device was explanted and replaced.Device will be returned.
|
| |
|
Search Alerts/Recalls
|