EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 8300AB
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code Dyspnea (1816)
Date of Event 09/14/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was learned through implant patient registry and investigation that a patient with a 25mm 8300ab aortic valve was explanted after an implant duration of 7 years, 7 months due to poor leaflet mobility causing stenosis and moderate to severe regurgitation.The patient presented with sob and chest pain.The patient was taken to the icu post procedure.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional 'customer complaint'.The information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated sections h6 (investigation findings, investigation conclusions).Leaflet immobility or restricted motion has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Leaflet immobility or restricted motion occurring post-procedurally is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely cause is patient factors, including chronic kidney disease (ckd), hyperlipidemia (hld) , diabetes mellitus (dm), and coronary artery disease (cad).The subject device is not available for evaluation as it was discarded.The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.H11: corrected data: corrected section h6 (type of investigation).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE VALVE SYSTEM
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key23294320
Report Number2015691-2025-08340
Device Sequence Number4797847
Product Code LWR
UDI-Device Identifier00690103194500
UDI-Public(01)00690103194500(17)210830
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P150036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number8300AB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2025
Supplement Date Received by Manufacturer11/05/2025
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date11/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient SexMale
-
-