• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. KL 1.8 FIBERTAK, SHOULDER; NON-DEGRADABLE FIXATN FASTENER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. KL 1.8 FIBERTAK, SHOULDER; NON-DEGRADABLE FIXATN FASTENER Back to Search Results
Model Number KL 1.8 FIBERTAK, SHOULDER
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a labral repair surgery the suture of the anchor tore.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 -complaint allegation is not confirmed.-visual inspection of the knotless 1.8 fibertak® soft anchor, gen2, with #2 suture, qty.1, identified that no anchor/sutures were returned for investigation.It was found that the fibertak® inserter assembly was fractured at the distal end of the shaft.-functional testing was not performed due to the missing anchor and sutures.No problem found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KL 1.8 FIBERTAK, SHOULDER
Common Device Name
NON-DEGRADABLE FIXATN FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key23294796
Report Number1220246-2025-04595
Device Sequence Number4797858
Product Code MBI
UDI-Device Identifier00888867362857
UDI-Public(01)00888867362857(11)250429(17)300331(10)15427713
Combination Product (Y/N)N
Initial Reporter CountryEZ
PMA/510(K) Number
K221396
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 01/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKL 1.8 FIBERTAK, SHOULDER
Device Catalogue NumberAR-3636-1
Device Lot Number15427713
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date10/06/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2025
Supplement Date Received by Manufacturer12/22/2025
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date01/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-