| Catalog Number |
20-650 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
11/11/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "grade i capsule" confirmed via mri and "redundant axillary elastosis".This record is for the left side.The device was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: visibility/palpability.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "grade i capsule" confirmed via mri and "redundant axillary elastosis".This record is for the left side.The device was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.¿ visibility/palpability: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (creases, wear abrasion and deformation) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|