| Model Number |
E141 |
| Medical Device Problem Codes |
High impedance (1291); Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited low intrinsic amplitude and high shock impedance out of range.The device remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited low intrinsic amplitude and high shock impedance out of range.The device remains in service.No adverse patient effects were reported.Additional information was obtained, the right ventricular (rv) lead is suspected to have broken and having issues, is currently in service and no adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Investigation of the available information determined this device exhibited oversensing of noise generate that is related to intermittent increases in impedance measurements.Additionally, engineering analysis and testing of returned products has identified that any repeated, small movements of the lead terminal ring can create wearing of the lead terminal ring and generate microscopic particles, which may accumulate and oxidize over time.This can impact the connection between the spring contact and the lead ring, resulting in intermittent changes in impedance measurements.Please refer to the description field for more information regarding the specific circumstances of this event.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited low intrinsic amplitude and high shock impedance out of range.The device remains in service.No adverse patient effects were reported.Additional information was obtained, the right ventricular (rv) lead is suspected to have broken and having issues, is currently in service and no adverse patient effects were reported.
|
| |
|