BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE SG; IMPLANTABLE LEAD Back to Search Results
Model Number 0180
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this right ventricular (rv) lead exhibited a gradual rise in shock impedance measurements, reaching values of 136 ohms.Technical services (ts) was consulted and continue to monitor was recommended.No adverse patient effects were reported.This lead remains in service.
 
Event or Problem Description
It was reported that this right ventricular (rv) lead exhibited a gradual rise in shock impedance measurements, reaching values of 136 ohms.Technical services (ts) was consulted and continue to monitor was recommended.No adverse patient effects were reported.This lead remains in service.Additional information was received indicating that the lead was surgically abandoned and replaced.No additional adverse patient effects were reported.
 
Event or Problem Description
It was reported that this right ventricular (rv) lead exhibited a gradual rise in shock impedance measurements, reaching values of 136 ohms.Technical services (ts) was consulted and continue to monitor was recommended.No adverse patient effects were reported.This lead remains in service.Additional information was received indicating that the lead was surgically abandoned and replaced.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific and physical analysis could not be conducted.Investigation of the available information/data determined that the observed gradual rise in lvsi measurements was likely associated with calcification, as there was no conclusive evidence of compromised physical or electrical integrity of the lead.This condition may develop over time and can be reflected in a gradual and sustained increase in shock impedance measurements.Boston scientific issued a field safety notice to physicians in july 2025 regarding the potential for reliance defibrillation leads with expanded polytetrafluoroethylene (eptfe) coating to exhibit a pattern of gradually increasing lvsi measurements due to calcification, which may result in elevated high voltage shock impedance (hvsi) measurements and reduced shock efficacy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE SG
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key23296462
Report Number2124215-2025-73204
Device Sequence Number11977793
Product Code LWS
UDI-Device Identifier00802526531149
UDI-Public(01)00802526531149(17)150816
Combination Product (Y/N)Y
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Remedial Action Notification,Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/16/2015
Device Model Number0180
Device Catalogue Number0180
Device Lot Number309908
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2025
Supplement Date Received by Manufacturer10/14/2025
02/23/2026
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date10/21/2025
02/25/2026
Date Device Manufactured08/16/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age74 YR
Patient SexFemale
-
-