Model Number 173050G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a laparoscopic nephrectomy with robot support, at the time of specimen removal, when collecting the specimen, there was an unusual sensation in the shaft, and the bag did not open properly.The internal shaft and string broke when the device was forced open.The bag was still used and the specimen was taken out.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the specimen pouch and string were fully attached to the metal ring.The specimen pouch was not received.The gold ring and string were disengaged from the plunger.The white shaft and ring were disengaged from the device.It was reported that there was an unusual sensation in the shaft, the bag did not open properly, and the internal shaft and string broke when the device was forced open.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the plunger is retracted after bag deployment before the bag is completely cinched by pulling the gold ring.This premature retraction causes undesirable bag detachment.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: if bag does not pull through easily, enlarge the incision to accommodate easy removal for specimens with diameters larger than the trocar site incision.Failure to follow this warning could cause the bag to break and the specimen to fall back into the body cavity.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CATCH GOLD
Common Device Name
LARYNGOSCOPE, ENDOSCOPE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key23298597
Report Number9612501-2025-02673
Device Sequence Number17340146
Product Code GCI
UDI-Device Identifier10884523000955
UDI-Public(01)10884523000955(17)280131(10)J5B2629MY
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K922123
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number173050G
Device Catalogue Number173050G
Device Lot NumberJ5B2629MY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2025
Supplement Date Received by Manufacturer11/18/2025
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date12/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-