| Catalog Number |
02K41-27 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Section a1 - patient identifier: complete sample id is (b)(6).An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated architect stat troponin-i and provided the following data: customer troponin cuttoff is 50 ug/l.On (b)(6) 2025, sid (b)(6) initial result was 74 ug/l and repeat result was 48 ug/l.Patient information: 63-year-old male.The customer confirmed that no patient care was impacted.Per the customer discrepant results were not reported out to the medical provider.There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, device history record review, field data review and labeling review.The data and information provided by the customer were reviewed and supported the complaint issue.A review of ticket trending data for the architect stat troponin-i did not identify an increase in complaint activity related to the complaint issue.Additionally, an increase in complaint activity was not identified for reagent lot 36403un25.A device history record review was performed on lot 36403un25, which did not show any potential non-conformances, deviations, or non-conformances.The overall performance of architect stat troponin-i reagents in the field was reviewed using data from customers worldwide.The patient median data (divided into 3 groups) and cal f rlu mean were analyzed and compared to an established control limit.This evaluation indicated that all three levels of the patient medians and the cal f rlu mean for the lot 36403un25 are within the established limits.Labeling was reviewed and found to adequately address the customer's issue.Based on the investigation, no systemic issue or deficiency of the architect stat troponin-i reagent lot 36403un25 was identified.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated architect stat troponin-i and provided the following data: customer troponin cuttoff is 50 ug/l on (b)(6) 2025, sid (b)(6), initial result was 74 ug/l and repeat result was 48 ug/l.Patient information: 63-year-old male the customer confirmed that no patient care was impacted.Per the customer discrepant results were not reported out to the medical provider.There was no reported impact to patient management.
|
| |
|