| Catalog Number |
00770100000 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation into the reported event has been initiated under (b)(6).Once the investigation has been completed, a follow up report will be submitted with the investigation results and any actions taken by the manufacturer.G2 country: australia.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, the device had a damaged grate under the cutter.There was no patient injury.Due diligence is complete and there is no additional information available.
|
| |
|
Event or Problem Description
|
|
There was no additional information associated with this malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).Following sections were updated or corrected: b4, b5, d4, d9, g3, g6, h2, h3, h4, h6 and h11.Review of the most recent repair record determined the comb was damaged.The comb was replaced to resolve the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to component failure as the device exhibits wear and tear from repeated use.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|