| Model Number |
0684-00-0498-01 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the intra-aortic balloon therapy, the inflation was poor and the patient's condition was critical.After immediate replacement, the inflation was good and the patient was out of danger.The operation was successful and there was no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to character restrictions, event site name: (b)(6).Device has not been returned to manufacturer, supplemental report will be submitted upon completion of additional findings.Complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields - b4, d9, g3, g6, h2, h3, h4, h6 (type of investigation, investigation findings, investigation conclusions), h11.No decon or visual inspection performed since product was not returned and could not be evaluated.Therefore, we were unable to confirm or determine a root cause for the reported failure.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: b4, d9, g3, g6, h2, h11.
|
| |
|