| Catalog Number |
0520161A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
10/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Event or Problem Description
|
|
A user facility¿s registered nurse (rn) reported that a visible internal dialyzer blood leak occurred immediately after the initiation of the patient¿s hemodialysis (hd) treatment.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were not used.Fresenius bloodlines were not in use.The patient¿s estimated blood loss (ebl) was approximately 150ml.There was no patient serious injury or medical intervention required as a result of this event.The patient did not finish the treatment.The complaint sample is not available to be returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.
|
| |
|
Event or Problem Description
|
|
A user facility¿s registered nurse (rn) reported that a visible internal dialyzer blood leak occurred immediately after the initiation of the patient¿s hemodialysis (hd) treatment.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were not used.Fresenius bloodlines were not in use.The patient¿s estimated blood loss (ebl) was approximately 150ml.There was no patient serious injury or medical intervention required as a result of this event.The patient did not finish the treatment.The complaint sample is not available to be returned to the manufacturer for evaluation.
|
| |
|