|
ETHICON INC. PRLNE BLU 36IN 4-0 D/A BB; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
|
Back to Search Results |
|
| Catalog Number |
8581H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 10/15/2025.H6 component code: g07002 - pending evaluation of returned device.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested the following was obtained: - when did the needle detach from the suture thread (removal from package / during handling prior to use on patient/ during passage through tissue / during tying / post-op)? - how many needles detach from the suture thread? - provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep).I personally do not have any additional information currently.The person i have been in contact with is dw a device has been received, however it has not yet been evaluated.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.The sales rep reported that during an unspecified procedure, the needle kept popping off the suture.No patient consequences were reported.The device is available.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Date sent to the fda: 11/18/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found additional information: d9, h3, h6.H3 investigational summary: the product was returned to ethicon for evaluation.One open box with twenty-eight representative unopened samples was received for analysis.Product code 8581h.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were evident on the exterior packaging.Following the sampling plan, a visual inspection was conducted on thirteen samples.The packets were opened, and swage and attachment areas appeared as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|