ZIMVIE US CORP LLC CERTAIN GOLD-TITE HEXED SCREW; DENTAL SCREW Back to Search Results
Catalog Number IUNIHG
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).
 
Event or Problem Description
It was reported that screw loosened.Psp will send replacement screw.
 
Additional Manufacturer Narrative
The following sections have been updated: b4: date of this report.D4: lot number.D4: udi number.D9: device availability.G3: date received by manufacturer.G6: checked "follow-up".H2: checked follow-up type.H10: added manufacturer narrative.
 
Event or Problem Description
No further event information available at the time of this report.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
Zimvie did not receive one (1) iunihg, (certain® gold-tite® hexed screw) for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Functional testing to recreate the reported event could not be performed due to the nature of the device & event.This complaint refers to the specific device being investigated for this complaint record.Dhr review was completed for the subject lot number 1293980.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1293980 for similar events and two (2) other complaints were identified (b)(4).The customer did not submit images for the reported event.Based on the investigation and risk management file review as per rmf rm-00057-haz rev.22, the most likely root cause determined from the investigation can be attribute to the re-tightened of the screw at the time of loosening although being a single-use device.Therefore, based on the available information, a device malfunction could not be verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report, g3: date received by manufacturer, g6: checked "follow-up" h6: entered evaluation codes, h10: added manufacturer narrative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN GOLD-TITE HEXED SCREW
Common Device Name
DENTAL SCREW
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key23301448
Report Number0001038806-2025-02430
Device Sequence Number15940835
Product Code NHA
UDI-Device Identifier00844868008538
UDI-Public(01)00844868008538(17)300122(10)1293980
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K072642
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberIUNIHG
Device Lot Number1293980
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date11/04/2025
11/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-