• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL SHERIDAN/HVT ET TUBE, HVT, 7.5; TUBE, TRACHEAL (W/WO CONNECTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL SHERIDAN/HVT ET TUBE, HVT, 7.5; TUBE, TRACHEAL (W/WO CONNECTOR Back to Search Results
Catalog Number 5-10315
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "patient had been recently intubated- respiratory therapy noted what they thought to be a leak from the ett cuff.Patient was reintubated by the provider.This rn injected air into original ett cuff and it was visibly deflating, not holding air as intended.No harm related to patient.".
 
Additional Manufacturer Narrative
(b)(4).The reported complaint of "leak - device leak - cuff" was confirmed based upon the investigation of the sample received.The customer returned one unit 5-10315 et tube, hvt, 7.5 for investigation.The returned et tube was found to have a hole in the balloon cuff which prevented it from being able to stay inflated.The ifu states, "the tracheal tube cuff inflation system, including the valve, should be checked prior to intubation.If a malfunction is detected in any part of the system, the tube should not be used." "care should be taken to avoid damaging the thin-walled cuffs during intubation.If cuff is damaged, the tube should not be used." the ifu also states, "cuff pressure should be monitored routinely to assure that an adequate tracheal seal is maintained and that the cuff has not become over-inflated.The tracheal tube cuff should be slowly inflated with the minimum amount of air required to provide an effective tracheal seal.Do not inflate with a measured volume of air or by "feel" of pressure in the syringe since little resistance should be felt during inflation.".A device history record review was performed on potential lot based on sales history, and no relevant findings were identified.All et tubes are 100% inspected for inflation and deflation at the time of manufacturing.It is unlikely that this type of damage was present at the time of manufacturing.Therefore, based upon the damage observed, unintentional user error caused or contributed to this event.
 
Event or Problem Description
It was reported that "patient had been recently intubated- respiratory therapy noted what they thought to be a leak from the ett cuff.Patient was reintubated by the provider.This rn injected air into original ett cuff and it was visibly deflating, not holding air as intended.No harm related to patient.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHERIDAN/HVT ET TUBE, HVT, 7.5
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key23301545
Report Number3003898360-2025-00768
Device Sequence Number9952975
Product Code BTR
UDI-Device Identifier04026704397044
UDI-Public(01)04026704397044(11)250308(17)300307(10)73C2500203(21)
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K822082
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5-10315
Device Lot Number73C2500203
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2025
Supplement Date Received by Manufacturer10/30/2025
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date10/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED.; NOT REPORTED.
Patient SexUnknown
-
-