| Catalog Number |
0165SI12 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter facility name: (b)(6).The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon ruptured of the foley catheter and was removed spontaneously.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed ¿ cause unknown.The reported failure was able to be reproduced.The product was used for urological care.Based on the evaluation, sac burst along lumen without missing piece was observed on the returned catheter.However, the exact cause of how and when the problem occurred could not be determined.A potential root cause could be user related (example: contact with sharp object)/ exposure to petrolatum based products/ mechanical failure/ thin sac.The labeling and instructions for use were reviewed and found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Initial reporter facility name: (b)(6).Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon ruptured of the foley catheter and was removed spontaneously.
|
| |
|