Catalog Number 0520161A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 10/07/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A user facility¿s registered nurse (rn) reported that a visible internal dialyzer blood leak occurred two hours and fifteen minutes after the initiation of the patient¿s hemodialysis (hd) treatment.The machine, a fresenius 2008t machine, did not alarm because the blood had not yet reached the blood leak detector before the leak was discovered.Blood leak test strips were not used.Fresenius bloodlines were not in use.The patient¿s estimated blood loss (ebl) was approximately 150ml.There was no patient serious injury or medical intervention required as a result of this event.The patient did not restart treatment.The complaint sample is not available.
 
Additional Manufacturer Narrative
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
 
Event or Problem Description
A user facility¿s registered nurse (rn) reported that a visible internal dialyzer blood leak occurred two hours and fifteen minutes after the initiation of the patient¿s hemodialysis (hd) treatment.The machine, a fresenius 2008t machine, did not alarm because the blood had not yet reached the blood leak detector before the leak was discovered.Blood leak test strips were not used.Fresenius bloodlines were not in use.The patient¿s estimated blood loss (ebl) was approximately 150ml.There was no patient serious injury or medical intervention required as a result of this event.The patient did not restart treatment.The complaint sample is not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOFLOW F4 LF 15/CS 0.7SA ETO
Common Device Name
DIALYZER, CAPILLARY, HOLLOW FIBER
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key23301796
Report Number0001713747-2025-00719
Device Sequence Number10027907
Product Code FJI
UDI-Device Identifier00840861100026
UDI-Public00840861100026
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K874872
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0520161A
Device Lot Number24PU04013
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2025
Supplement Date Received by Manufacturer12/21/2025
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date01/05/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/17/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS HEMODIALYSIS MACHINE; FRESENIUS HEMODIALYSIS MACHINE
Patient SexUnknown
-
-