• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION R SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION R SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number 30001004101130012
Medical Device Problem Code Defibrillation/Stimulation Problem (1573)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was evaluated by zoll medical corporation.The customer's report was duplicated and attributed to a relay on the pace/defib engine board.The pace/defib engine board will be replaced to resolve the report.The board was scrapped after testing.The device will be recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Event or Problem Description
Complainant alleged that during a routine shift check by a clinician, the device displayed a "defib fault 76" message.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key23301967
Report Number1220908-2025-03722
Device Sequence Number11984459
Product Code MKJ
UDI-Device Identifier00847946017453
UDI-Public(01)00847946017453
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P160022
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number30001004101130012
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date11/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-