This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history: the device lot number is unknown; therefore, a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that on (b)(6) 2025, the patient underwent a tka surgery for oa with the products in question.The surgery was completed successfully with no surgical delay.After about two weeks of non-weight bearing after surgery, rehabilitation began, but the patient complained of pain, so photographs were taken to check.Subsequent follow-up observation revealed that the fracture had widened and the tibia had become tilted.The surgeon commented "when the tibial base plate was placed posteriorly and cementless, the pegs interfered, causing a spur after surgery.It is believed that the spur then widened because rehabilitation began before bone fusion had begun." no further information is available.
|