| Catalog Number |
FM-410500 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/25/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture", "removal from textured to smooth due to the concerns of the product".This record is for the right side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture", "removal from textured to smooth due to the concerns of the product".This record is for the right side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: observed broken device through microscopic inspection assessed as fold flaw opening and missing piece of shell assessed as inconclusive.No further actions are required as it is not related to the manufacturing process.¿ anxiety-product/procedure: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (creases and wear abrasion) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture", "removal from textured to smooth due to the concerns of the product".This record is for the right side.Device was explanted.
|
| |
|
Search Alerts/Recalls
|